Product Details
Outstanding Crossability in Tight Lesions1
The proprietary balloon material for small sizes enables slim balloon shoulders while maintaining durability. Coupled with hydrophilic balloon coating, Pantera Pro excels in tight lesions.
43% Less Friction2 during Kissing Balloon Technique
The reduced distal shaft profile lowers friction when using two balloons in a 6 F guiding catheter.
Any combination of two diameters not larger than 3.5 mm within a 6F guiding catheter with a minimal inner diameter of 0.070"/1.78 mm.
38% More Push3 to Reach Target Lesion
BIOTRONIK’s unique Enhanced Force Transmission (EFT) shaft results in excellent pushability by connecting the proximal hypotube directly to the distal guide wire support section.
Technical Data
Proximal Shaft | |
---|---|
Design |
Hypotube design |
Diameter |
2.0F |
Shaft markers | 92 cm and 102 cm from tip |
Distal Shaft | |
---|---|
Guiding catheter | 5F (min. I.D. 0.056" / 1.42mm) |
Guide wire diameter | 0.014" |
Lesion entry profile | 0.017" |
Usable length | 140 cm |
Balloon material | Semi-Crystalline Co-Polymer |
Balloon folding | ø 1.25 - 1.5 mm: Two-fold; ø 2.0 - 4.0 mm: Tri-fold |
Balloon markers | Platinum-iridium: ø 1.25 - 1.5 mm one mid-marker; ø 2.0 - 4.0 mm two markers |
Coating distal shaft | Hydrophilic (end of balloon to guide wire exit port) |
Balloon and tip coating | ø 1.25 - 2.0 mm: Hydrophilic ø 2.50 - 4.0 mm: Hydrophobic |
Kissing balloon technique | 6 F guiding catheter (min. I.D. 0.070"/1.78 mm), up to ø 3.5 mm |
Diameter | 2.6 F (ø 1.25 - 2.0 mm); 2.7 F (ø 2.5 - 3.5 mm); 2.9 F (ø 4.0 mm) |
Compliance Chart
Balloon Diameter x Length (mm) | |||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Nominal Pressure | atm4 | 7 | 7 | 7 | 7 | 7 | 7 | 7 | |||||||||||
(NP) | ø (mm) | 1.24 | 1.49 | 2.01 | 2.49 | 3.08 | 3.62 | 3.95 | |||||||||||
Rated Burst Pressure | atm4 | 14 | 14 | 14 | 14 | 14 | 14 | 14 | |||||||||||
(RBP) | ø (mm) | 1.37 | 1.72 | 2.23 | 2.93 | 3.5 | 4.06 | 4.55 |
Ordering Information
Contact
1 1.25 - 2.0 mm diameter, when compared to key competitors. BIOTRONIK data on file.
2 vs Trek (Abbott), BIOTRONIK data on file.
3 vs Mini Trek (Abbott), BIOTRONIK data on file.
4 1 atm = 1.013 bar